Table 1 The result of detection rate, false positives, and number of undetected lesions according to the different threshold values

From: Computer-aided detection of breast lesions in DCE-MRI using region growing based on fuzzy C-means clustering and vesselness filter

Threshold 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Detection rate 0.98 0.95 0.94 0.92 0.89 0.86 0.81 0.74 0.59 0.52 0
False positives per case 13 5.91 5.29 4.41 3.53 2.80 2.20 1.40 0.72 0.37 0
Number of undetected lesions 3 8 10 13 18 23 32 44 69 81 170